Trolex AirXD Power Connections

Modified on Mon, 19 Feb at 1:29 PM

The Air XD can be ordered and configured to allow connection to a standard 100 to 240 V ac power supply or a 9 to 36 V dc power supply. Before connecting a power supply to the instrument, ensure that the supply source is compatible with the instrument and information outlined on the appropriate rating plate. See below for rating plate details:



AC Power Supply

For instruments connected to an ac power supply, it is the responsibility of the installer to ensure that the instrument is installed with the following protection:

- An external fuse or circuit breaker at a maximum of 5 A.

- Externally earthed.


See label below for details.


For instruments connected to an ac power supply, ensure that the equipment is protected by an external fuse or circuit breaker at a maximum of 5 A.



DC Power Supply

For instruments connected to a dc power supply, it is the responsibility of the installer to ensure that the instrument is installed with a dc supply, meeting re-enforced insulation requirements of EN61010-1 or equivalent.